Reported Aug 22, 2018 · Terminated

NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A) Product Usage: The NxStage Express Fluid…

NxStage Medical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: NxStage Medical, Inc. · 350 Merrimack St · Lawrence, MA · 01843-1748

Is this the product you have?

    • NxStage Express Fluid Warmer (FW-300
    • FW-300-A
    • FW-300-WW
    • FW-301
  • FW-301-A) Product UsageThe NxStage Express Fluid Warmer is an accessory to the NxStage System One used to warm therapy fluids prior to administration
Lot codes and dates
  • All units are affected. As of 6/11/18, the last serial number manufactured was W35452.
Amount recalled
14975

Where it was sold

  • Nationwide

Why it was recalled

There is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is plugged in. The firm has received 7 reports of smoke, sparks, or flames coming from the Express Fluid Warmer since 2016. No adverse events have been reported.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new NxStage Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 28, 2020 · Data as of Sep 19, 4:48 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.