Reported Aug 19, 2020 · Terminated

uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified…

Shanghai United Imaging Healthcare Co., Ltd.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Shanghai United Imaging Healthcare Co., Ltd. · No. 2258 Chengbei Rd, Jiading Ind. · Shanghai, N/A · N/A

Is this the product you have?

  • uEXPLORER PET/CT Diagnostic
  • Imaging System - Product UsageThis system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology
Lot codes and dates
    • Serial No. F00002
    • GTIN 06971576832040
Amount recalled
1 unit distributed to Texas

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide
  • TX

Why it was recalled

1) 1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Shanghai United Imaging Healthcare Co., Ltd. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 7, 2022 · Data as of Sep 19, 6:48 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.