Reported Sep 30, 2015 · Terminated
The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica ASP6025 Tissue Processor is an…
Leica Microsystems, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Leica Microsystems, Inc. · 1700 Leider Ln · Buffalo Grove, IL · 60089-6622
Is this the product you have?
- The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica ASP6025 Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.
- Lot codes and dates
- Product CodeLeica
- ASP6025 Serial Numbers
163169200-729731733735737739741743745747749751753755757759761763765767769771773775777779781783785787789791793795797799801803805807809811
- Amount recalled
- U.S. Quantity Distributed: 163 systems; Foreign Quantity Diistributed: 410 systems.
Where it was sold
Why it was recalled
The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Leica Microsystems, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.