Reported Aug 26, 2020 · Terminated
Phillips Module Charger, Product #: 989803191031 - Product Usage: This Module Charger is used in the Expression MR400…
Philips North America, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips North America, LLC · 3000 Minuteman Rd · Andover, MA · 01810-1032
Is this the product you have?
- Phillips Module Charger, Product #: 989803191031 -
- Product UsageThis Module Charger is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner
- Lot codes and dates
- Product #: 989803191031
- Manufacturing date codesprior to 2020-03
- Amount recalled
- 9,705 units
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Why it was recalled
A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips North America, LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.