Reported Aug 22, 2018 · Terminated

Sorin Platinium CRT-D 1741 DF-4, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy…

Sorin CRM SAS

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Sorin CRM SAS · Cvk · Clamart Cedex · N/A

Is this the product you have?

  • Sorin Platinium CRT-D 1741 DF-4, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator
Lot codes and dates
  • Model TDF037U
  • UDI GTIN 08031527015484 Serial Numbers
    • 637DC0D0
    • 637DC13F
    • 625DC128
    • 622DC17A
    • 620DC034
    • 627DC0E0
    • 619DC101
    • 620DC07F
    • 622DC02C
    • 619DC0E1
Amount recalled
10

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrillation shock therapy.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Sorin CRM SAS recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 6, 2024 · Data as of Sep 19, 5:27 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.