Reported Oct 7, 2015 · Terminated

Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System Product Usage: Stretcher for pre-hospital…

Ferno-Washington Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ferno-Washington Inc · 70 Weil Way · Wilmington, OH · 45177-9371

Is this the product you have?

  • Ferno iNX Wheeled Stretcher, Integrated Patient Transport &
  • Loading System Product UsageStretcher for pre-hospital patient transport
Lot codes and dates
  • 102 of 102 codes

    • Serial #s: 15F001703
    • 15F001704
    • 15F001706
    • 15F001707
    • 15F001708
    • 15F001709
    • 15F001710
    • 15F001711
    • 15F001712
    • 15F001713
    • 15F001715
    • 15F001717
    • 15F001719
    • 15F001721
    • 15F001722
    • 15F001723
    • 15F001724
    • 15F001725
    • 15F001726
    • 15F001727
    • 15F001728
    • 15F001729
    • 15F001730
    • 15F001731
    • 15F001732
    • 15F001733
    • 15F001734
    • 15F001735
    • 15F001736
    • 15F001737
    • 15F001738
    • 15F001739
    • 15F001740
    • 15F001741
    • 15F001742
    • 15F001744
    • 15F001745
    • 15F001746
    • 15F001747
    • 15F001748
    • 15F001749
    • 15F001750
    • 15F001751
    • 15F001752
    • 15F001753
    • 15F001754
    • 15F001755
    • 15F001756
    • 15F001757
    • 15F001758
    • 15F001759
    • 15F001760
    • 15F001761
    • 15F001762
    • 15F001763
    • 15F001764
    • 15F001765
    • 15F001766
    • 15F001767
    • 15F001768
Amount recalled
72 units

Where it was sold

Why it was recalled

Hole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ferno-Washington Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jan 20, 2016 · Data as of Sep 19, 3:23 PM UTC