Reported Sep 28, 2016 · Terminated

CADstream software. Product Usage: CADstream is an image processing system designed to assist in the visualization,…

Merge Healthcare, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Merge Healthcare, Inc. · 900 Walnut Ridge Dr · Hartland, WI · 53029-8347

Is this the product you have?

  • CADstream software
  • Product UsageCADstream is an image processing system designed to assist in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. CADstream also is intended to provide workflow efficiency and interventional planning tools
Lot codes and dates
  • All versions - Versions 4.1.3 and prior versions
Amount recalled
1,349 sites potentially have the affected versions

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

An incorrect biopsy or missed target could result if the incorrect grid is selected within the application.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Merge Healthcare, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 5, 2020 · Data as of Sep 19, 5:20 PM UTC