Reported Aug 12, 2026 · Ongoing
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph
GE Medical Systems Information Technologies Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Medical Systems Information Technologies Inc · 9900 W Innovation Dr · Milwaukee, WI · 53226-4856
Is this the product you have?
GE Healthcare MAC VU360REF# part number 2030361-023Electrocardiograph
- Lot codes and dates
- GTIN 00198953107263
- Serial Numbers
163 of 163 codes
VU221150078SAVU222400121SAVU223230048SAVU222520132SAVU219400033SAVU219400034SAVU219400045SAVU219400055SAVU220270027SAVU220510048SAVU220510078SAVU221100161SAVU221100181SAVU221150116SAVU222120212SAVU222120220SAVU222120239SAVU222120243SAVU222120249SAVU222120250SAVU222120253SAVU222120258SAVU222120260SAVU222120261SAVU222120266SAVU222120269SAVU222140004SAVU222200296SAVU222520114SAVU223120014SAVU223120041SAVU223420119SAVU224050158SAVU224070091SAVU224200028SAVU224420159SAVU224520140SAVU225170260SAVU225420050SAVU225520037SAVU226040025SAVU419320042SAVU421490135SAVU421490151SAVU421490153SAVU421490155SAVU421490159SAVU422460150SAVU422460194SAVU422460203SAVU422460303SAVU422480164SAVU422500239SAVU422500240SAVU422500241SAVU422500242SAVU422500244SAVU422500245SAVU422500246SAVU422500247SA
- Amount recalled
- 163 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new GE Medical Systems Information Technologies Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.