Reported Oct 5, 2016 · Terminated
R&D Systems, Quantikine¿ IVD¿sTfR Immunoassay Kit, for the quantitative determination of soluble transferrin receptor…
R & D Systems, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: R & D Systems, Inc. · 614 McKinley Pl NE · Minneapolis, MN · 55413-2610
Is this the product you have?
- R&D Systems, Quantikine¿ IVD¿sTfR Immunoassay Kit, for the quantitative determination of soluble transferrin receptor (sTfR) concentrations in human serum and plasma For in Vitro Diagnostic Use, REF DTFR1.
- Lot codes and dates
331636331637332060332090332924334184335062335740336694339307340362
- Amount recalled
- 725 kits (502 US, 223 OUS)
Where it was sold
Why it was recalled
R&D Systems, Inc. received two customer complaints that results of the Controls provided with the Quantikine¿ IVD¿ Human sTfR Immunoassay kit were out of the range high. Internal testing within R&D Systems confirmed a problem with Control results being out of range high.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new R & D Systems, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.