Reported Sep 19, 2018 · Terminated
Portable Suction Pump, (a) Style 758010, Model CARE-E-VAC 3, Ohio Medical LLC; (b) Style 758000, Model CARE-E-VAC 3,…
Ohio Medical Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ohio Medical Corporation · 1111 Lakeside Dr · Gurnee, IL · 60031-2489
Is this the product you have?
- Portable Suction Pump, (a) Style 758010, Model CARE-E-VAC 3, Ohio Medical LLC; (b) Style 758000, Model CARE-E-VAC 3, Ohio Medical LLC The care-e-(R)3 is a portable medical aspirator used to suction fluids or foreign bodies from a patient. The primary intended use is to aspirate saliva, mucous, vomitus or other aspirant from the mouth or airway to allow adequate respiration or ventilation of the patient.
- Lot codes and dates
- (a)
- style 758010 Serial Numbers(b)
163 of 163 codes
CDS5000CDS5001CDS5002CDS5003CDS5004CDS5005CDS5006CDS5007CDS5008CDS5009CDS5010CDS5011CDS5012CDS5013CDS5014CDS5015CDS5016CDS5017CDS5018CDS5019CDS5020CDS5021CDS5022CDS5023CDS5024CDS5025CDS5026CDS5027CDS5028CDS5029CDS5030CDS5031CDS5032CDS5033CDS5034CDS5035CDS5036CDS5037CDS5038CDS5039CDS5040CDS5041CDS5042CDS5043CDS5044CEV3X0076CEV3X0077CEV3X0078CEV3X0079CEV3X0080CEV3X0081CEV3X0082CEV3X0083CEV3X0084CEV3X0085CEV3X0086CEV3X0087CEV3X0088CEV3X0089CEV3X0090
- style 758000 Serial Numbers
1,352 of 1,352 codes
CDS0081CDS0082CDS0083CDS0084CDS0085CDS0086CDS0087CDS0088CDS0089CDS0090CDS0091CDS0092CDS0093CDS0094CDS0095CDS0096CDS0097CDS0098CDS0099CDS0100CDS0101CDS0102CDS0103CDS0104CDS0105CDS0106CDS0107CDS0108CDS0109CDS0110CDS0111CDS0112CDS0113CDS0114CDS0115CDS0116CDS0117CDS0118CDS0119CDS0120CDS0121CDS0122CDS0123CDS0124CDS0125CDS0126CDS0127CDS0128CDS0129CDS0130CDS0131CDS0132CDS0133CDS0134CDS0135CDS0136CDS0137CDS0138CDS0139CDS0140
- Amount recalled
- 1443 units
Where it was sold
- Nationwide
Why it was recalled
The care-e-vac(R)3 Battery/AC Powered Portable Aspirator battery terminal may come in contact with the pump causing an electric short leading to damage to wiring and resulting in loss of function or an inoperable unit under battery power.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Ohio Medical Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.