Reported Sep 19, 2018 · Terminated
Edwards Lifesciences Thin-Flex Single Stage Venous Drainage Cannula, 38FRX40CM, resterilized.
SPS Sterilization, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: SPS Sterilization, Inc · 1110 N Gerald St · Nixa, MO · 65714-7445
Is this the product you have?
- Edwards Lifesciences Thin-Flex Single Stage Venous Drainage Cannula, 38FRX40CM, resterilized.
- Lot codes and dates
Code Number TF038L-RLot Number 59634935Serial Numbers 18302172017-0118302172017-0218302172017-0318302172017-0418302172017-0518302172017-0618302172017-0718302172017-0818302172017-0918302172017-10
- Amount recalled
- 10 units total
This recall covers 15 products
The agency filed one notice per product. You are reading one of them; the other 14 are below.
2 more are in the archive.
Where it was sold
Why it was recalled
The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new SPS Sterilization, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.