Reported Sep 19, 2018 · Terminated

VITROS Chemistry Products Cl- Slides, Product Code 684 4471 Product Usage For in vitro diagnostic use only. VITROS…

Ortho-Clinical Diagnostics

Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ortho-Clinical Diagnostics · 100 Indigo Creek Dr · Rochester, NY · 14626-5101

Is this the product you have?

  • VITROS Chemistry Products Cl- Slides, Product Code 684 4471 Product Usage For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum, plasma and urine using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis
  • Product Usage
Lot codes and dates
  • GEN 10 and above
Amount recalled
420105

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride.

This product is unlikely to cause harm, but it breaks a safety or labelling rule.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ortho-Clinical Diagnostics recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 11, 2020 · Data as of Sep 19, 6:04 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.