Reported Oct 7, 2020 · Terminated

ToxCup Drug Screen Cup (AMT Item Number/ AMT Description): HT29A/"ToxCup¿ Drug Screen Cup…

Ameditech Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ameditech Inc · 9940 Mesa Rim Rd · San Diego, CA · 92121-2910

Is this the product you have?

23 of 23 products

Products covered by this recall
#Product as published
1ToxCup Drug Screen Cup (AMT Item Number/ AMT Description): HT29A/"ToxCup¿ Drug Screen Cup COC300/MET500/OPI300/THC50/AMP1000/PCP25/OXY100 Adulteration Screen for CR - NI - PH - BL - S.G.", DT14/ "ToxCup¿ Drug Screen Cup THC50/COC150
2OPI300/MET500
3AMP500/BZO300
4BAR300/MTD300
5BUPG10/TCA1000
6MDMA500/OXY100
7PCP25/PPX300", HT15/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000", HT15A/ ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000 Adulteration Screen for CR - NI - PH - BL - S.G.", PT15/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/OPI300/THC50" DT13A/ "ToxCup¿ Drug Screen Cup THC50/COC150
8OPI300/MET500
9AMP500/BZO300
10BAR300/MTD300
11BUPG10/TCA1000
12MDMA500/OXY100 w/ Cr, Ni, pH, BI, S.G.", DT14A "ToxCup¿ Drug Screen Cup THC50/COC150
13OPI300/MET500
14AMP500/BZO300
15BAR300/MTD300
16BUPG10/TCA1000
17MDMA500/OXY100
18PCP25/PPX300 w/ Cr, Ni, pH, Bl, S.G. Finished Devices", DT12A/"ToxCup¿ Drug Screen Cup THC50/COC150
19OPI300/MET500
20AMP500/BZO300
Lot codes and dates
  • ALL Lots Within Expiry
Amount recalled
189,175 units

This recall covers 23 products

The agency filed one notice per product. You are reading one of them; the other 22 are below.

10 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ameditech Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 30, 2024 · Data as of Sep 19, 6:11 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.