Reported Sep 6, 2017 · Terminated

MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF…

Encore Medical, LP

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Encore Medical, Lp · 9800 Metric Blvd · Austin, TX · 78758-5445

Is this the product you have?

  • MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN
  • Catalog Numbers
    • REF S-200956 - 5mm PIN
    • SIZE 46 REF S-200957 - 5mm PIN
    • SIZE 50 REF S-200958 - 5mm PIN
    • SIZE 58
Lot codes and dates
    • Catalog number
    • Lot Number S-200956
    • 535U1000 S-200957
    • 536U1000 S-200958
    • 537U1000
Amount recalled
5 units

Where it was sold

Why it was recalled

It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Encore Medical, LP recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 28, 2023 · Data as of Sep 19, 6:03 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.