Reported Sep 25, 2024 · Ongoing

STREAMLINE LONG-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis.…

B Braun Medical Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: B Braun Medical Inc · 824 12th Ave · Bethlehem, PA · 18018-3524

Is this the product you have?

  • STREAMLINE LONG-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis
  • Model/Catalog NumberSL-2000M2095W
Lot codes and dates
  • UDI-DI 04046955348800; All Unexpired Lot Numbers
Amount recalled
142336 units

This recall covers 7 products

The agency filed one notice per product. You are reading one of them; the other 6 are below.

Where it was sold

  • Nationwide

Why it was recalled

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new B Braun Medical Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:42 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.