Reported Sep 26, 2018 · Terminated

VITROS Chemistry Products CREA Slides, Product Code 680 2584 Product Usage: VITROS CREA Slides (Product Code 680 2584)…

Ortho-Clinical Diagnostics

Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ortho-Clinical Diagnostics · 100 Indigo Creek Dr · Rochester, NY · 14626-5101

Is this the product you have?

  • VITROS Chemistry Products CREA Slides
  • Product Code 680 2584 Product UsageVITROS CREA Slides (Product Code 680 2584) - For in vitro diagnostic use only. VITROS Chemistry Products CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System
Lot codes and dates
    • GEN 7
    • GEN 10
    • GEN 12
    • GEN 13
    • GEN 15
    • GEN 16
    • GEN 18
Amount recalled
627543

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

The slides did not meet the current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for serum and urine samples and therefore affecting the low end of the measuring range for this assay. The current low end measuring range is 0.05mg/dL for serum and 1.2mg/dL for urine. Based on the CREA testing results for the LoQ, the revised low end CREA measuring range is 0.15mg/dL for serum and 3.2mg/dL for urine.

This product is unlikely to cause harm, but it breaks a safety or labelling rule.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ortho-Clinical Diagnostics recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.