Reported Oct 2, 2024 · Ongoing

Ingenia Elition S -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra…

Philips North America LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips North America Llc · 222 Jacobs St · Cambridge, MA · 02141-2289

Is this the product you have?

  • Ingenia Elition S -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF)
  • Numbers1) 782150 2) 782106
Lot codes and dates
  • Model UDI-DI
    • 782150 (01)00884838098329(21)
    • 782106 (01)00884838098329(21)
  • Serial Numbers
    • 26020
    • 26023
    • 26026
    • 26034
    • 46240
    • 46270
    • 46310
    • 46315
    • 46324
    • 46352
Amount recalled
10 units

This recall covers 13 products

The agency filed one notice per product. You are reading one of them; the other 12 are below.

Where it was sold

  • Nationwide

Why it was recalled

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips North America LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:32 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.