Reported Oct 4, 2017 · Terminated
Percuvance Scissors Tool Tip The Percutaneous Surgical System with 5mm attachments is indicated for the means to…
Teleflex Medical
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Teleflex Medical · 3015 Carrington Mill Blvd · Morrisville, NC · 27560-5437
Is this the product you have?
- Percuvance Scissors Tool Tip The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.
- Lot codes and dates
Lot #'s: 73A160060273C160017073D160048373D160062373E160017473F160014973F160020673G160039073H160000673H160035773H160050673K160047173L1600034and 73M1600327
- Amount recalled
- 20,039 total (US and OUS)
This recall covers 7 products
The agency filed one notice per product. You are reading one of them; the other 6 are below.
Where it was sold
Why it was recalled
For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Teleflex Medical recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.