Reported Sep 30, 2026 · Ongoing · Updated Oct 6, 2026

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

Biomerieux Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Biomerieux Inc · 595 Anglum Rd · Hazelwood, MO · 63042-2320

Is this the product you have?

  • VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811
Lot codes and dates
    • UDI-DI 03573026637712
    • Lot Number 1333393504
Amount recalled
97680 units

Why it was recalled

The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new produce recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Nov 4, 4:04 AM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.