Reported Oct 3, 2018 · Terminated
Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog…
Terumo Cardiovascular Systems Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Terumo Cardiovascular Systems Corporation · 6200 Jackson Rd · Ann Arbor, MI · 48103-9586
Is this the product you have?
- Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801764
- Lot codes and dates
- Serial Numbers
1522152315241525152815301526152715321544
- Spare Parts870509
- ASSEMBLY POWER MANAGER POD APS S/N
0309703099031000310103113031140311503143031550316703264
- Serial Numbers
- Amount recalled
- 10 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
Screws on a Terumo¿ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Terumo Cardiovascular Systems Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.