Reported Sep 30, 2026 · Ongoing · Updated Oct 24, 2026

Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella…

Abiomed, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Abiomed, Inc. · 22 Cherry Hill Dr · Danvers, MA · 01923-2575

Is this the product you have?

7 of 7 products

Products covered by this recall
Model / numberProduct as published
0042-0000-CA.Impella Controller, Packaged, CA
0042-0000-EU.Impella Controller, Packaged, EU
0042-0000-UK.Impella Controller, Packaged, UK
0042-0000-US.Impella Controller, Packaged, US
0042-0010-EU.Impella Optical Controller, Packaged, EU
0042-0010-US.Impella Optical Controller, Packaged, US
0042-0040-US.Optical, AIC, Impella Connect, Pkgd, US
Lot codes and dates
  • 1
  • Product Code0042-0000-CA
  • GTIN00813502011272
  • Serial NumberIC2687. 2
  • Product Code0042-0000-EU
  • GTIN00813502011289
  • Serial Numbers

    343 of 343 codes

    • IC3875
    • IC2380
    • IC2331
    • IC1721
    • IC1593
    • IC1587
    • IC4911
    • IC4231
    • IC4232
    • IC6844
    • IC2221
    • IC4628
    • IC6474
    • IC4502
    • IC1398
    • IC4055
    • IC3816
    • IC6810
    • IC6812
    • IC6449
    • IC6469
    • IC6441
    • IC6443
    • IC6440
    • IC6436
    • IC6437
    • IC6439
    • IC6272
    • IC6149
    • IC6081
    • IC5681
    • IC5682
    • IC5517
    • IC5177
    • IC5169
    • IC5086
    • IC4892
    • IC4852
    • IC4848
    • IC4875
    • IC4851
    • IC4869
    • IC4823
    • IC4824
    • IC4821
    • IC4822
    • IC4845
    • IC4840
    • IC4819
    • IC4820
    • IC4817
    • IC4818
    • IC4815
    • IC4816
    • IC4796
    • IC4797
    • IC4793
    • IC4795
    • IC4792
    • IC4794
  • Product Code0042-0000-UK
  • GTIN00813502011296
  • Serial Numbers
    • IC1494
    • IC3785
    • IC3784. 4
  • Product Code0042-0000-US
  • GTIN00813502010022
  • Serial Numbers

    2,690 of 2,690 codes

    • IC7027
    • IC7030
    • IC7032
    • IC6931
    • IC6930
    • IC6932
    • IC6929
    • IC6928
    • IC6923
    • IC6919
    • IC6920
    • IC6921
    • IC6922
    • IC6898
    • IC6897
    • IC6917
    • IC6916
    • IC6889
    • IC6894
    • IC6895
    • IC6893
    • IC6901
    • IC6909
    • IC6911
    • IC6910
    • IC6912
    • IC6908
    • IC6913
    • IC6896
    • IC6899
    • IC6900
    • IC6842
    • IC6840
    • IC6836
    • IC6829
    • IC6831
    • IC6832
    • IC6811
    • IC6774
    • IC6864
    • IC6830
    • IC6801
    • IC6800
    • IC6798
    • IC6799
    • IC6847
    • IC6771
    • IC6772
    • IC6850
    • IC6851
    • IC6852
    • IC6854
    • IC6822
    • IC6820
    • IC6819
    • IC6818
    • IC6767
    • IC6754
    • IC6743
    • IC6729
  • Product Code0042-0010-EU
  • GTIN00813502010961
  • Serial Numbers

    266 of 266 codes

    • IC6118
    • IC6119
    • IC5129
    • IC5042
    • IC5919
    • IC5282
    • IC6263
    • IC6261
    • IC5580
    • IC5592
    • IC5038
    • IC7156
    • IC6779
    • IC6776
    • IC6462
    • IC6418
    • IC6379
    • IC6416
    • IC6417
    • IC6383
    • IC6384
    • IC6377
    • IC6378
    • IC6344
    • IC6305
    • IC6336
    • IC6337
    • IC6304
    • IC6303
    • IC6342
    • IC6343
    • IC6302
    • IC6297
    • IC6298
    • IC6299
    • IC6282
    • IC6283
    • IC6278
    • IC6279
    • IC6277
    • IC6260
    • IC6259
    • IC6257
    • IC6258
    • IC6235
    • IC6256
    • IC6070
    • IC6231
    • IC6232
    • IC6234
    • IC6175
    • IC6226
    • IC6168
    • IC6169
    • IC6166
    • IC6167
    • IC6122
    • IC6123
    • IC6121
    • IC6174
  • Product Code0042-0010-US
  • GTIN00813502010985
  • Serial Numbers
    • IC7251
    • IC7037
    • IC7023
    • IC7021
    • IC6996
    • IC6998
    • IC6999
    • IC6975
    • IC6997
    • IC6974
    • IC6972
    • IC6963
    • IC6964
    • IC6959
    • IC6958
    • IC6957
    • IC5971
    • IC5970. 7
  • Product Code0042-0040-US
  • GTIN00813502011401
  • Serial NumbersIC7186
Amount recalled
3,322 units

Where it was sold

  • Nationwide

Why it was recalled

Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Abiomed, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Nov 4, 12:22 AM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.