Reported Sep 30, 2026 · Ongoing · Updated Oct 6, 2026

Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product…

Baxter Healthcare Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Baxter Healthcare Corporation · 1 Baxter Pkwy · Deerfield, IL · 60015-4625

Is this the product you have?

  • Brand NameVOLARA
  • System Product NameVOLARA, HC
  • NA A Model/Catalog NumberM08594A
  • Software VersionN/A
  • Product DescriptionVOLARA, HOMECARE
  • NORTH AMERICA ComponentNo
Lot codes and dates
  • Lot CodeUDI: 00887761981690
  • Serial NumbersW078SY3553 and Z309SY7853
Amount recalled
2

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

Why it was recalled

Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Baxter Healthcare Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Nov 4, 1:07 AM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.