Reported Oct 10, 2018 · Terminated
Bronchovideoscope BF-240
Olympus Corporation of the Americas
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Olympus Corporation of the Americas · 3500 Corporate Pkwy · Center Valley, PA · 18034-0610
Is this the product you have?
- Bronchovideoscope BF-240
- Lot codes and dates
- all lot codes
- Amount recalled
- 1467
This recall covers 30 products
The agency filed one notice per product. You are reading one of them; the other 29 are below.
17 more are in the archive.
Where it was sold
- Nationwide
Why it was recalled
The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
This product could cause temporary or reversible health problems — stop using it.