Reported Nov 4, 2026 · Ongoing
Philips Azurion 5 M20. Model Numbers: 722228, 722281.
Philips Medical Systems Nederland B.v.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Veenpluis 6 · Best
Is this the product you have?
- Philips Azurion 5 M20
- Model Numbers
722228722281
- Lot codes and dates
- 1
- Model Number722228
- UDI Numbers
253 of 253 codes
(01)00884838099234(21)49(01)00884838099234(21)159(01)00884838099234(21)180(01)00884838099234(21)273(01)00884838099234(21)132(01)00884838099234(21)533(01)00884838099234(21)47(01)00884838099234(21)522(01)00884838099234(21)32(01)00884838099234(21)23(01)00884838099234(21)25(01)00884838099234(21)62(01)00884838099234(21)48(01)00884838099234(21)153(01)00884838099234(21)431(01)00884838099234(21)113(01)00884838099234(21)529(01)00884838099234(21)659(01)00884838099234(21)145(01)00884838099234(21)567(01)00884838099234(21)539(01)00884838099234(21)645(01)00884838099234(21)308(01)00884838099234(21)333(01)00884838099234(21)335(01)00884838099234(21)570(01)00884838099234(21)560(01)00884838099234(21)44(01)00884838099234(21)569(01)00884838099234(21)418(01)00884838099234(21)280(01)00884838099234(21)616(01)00884838099234(21)568(01)00884838099234(21)471(01)00884838099234(21)428(01)00884838099234(21)465(01)00884838099234(21)102(01)00884838099234(21)176(01)00884838099234(21)103(01)00884838099234(21)178(01)00884838099234(21)109(01)00884838099234(21)16(01)00884838099234(21)234(01)00884838099234(21)566(01)00884838099234(21)571(01)00884838099234(21)463(01)00884838099234(21)647(01)00884838099234(21)317(01)00884838099234(21)562(01)00884838099234(21)552(01)00884838099234(21)680(01)00884838099234(21)724(01)00884838099234(21)561(01)00884838099234(21)517(01)00884838099234(21)203(01)00884838099234(21)773(01)00884838099234(21)260(01)00884838099234(21)107(01)00884838099234(21)100(01)00884838099234(21)108
- Serial Numbers
253 of 253 codes
491591802731325334752232232562481534311135296591455675396453083333355705604456941828061656847142846510217610317810916234566571463647317562552680724561517203773260107100108
- Model Number722281
- UDI Numbers
212 of 212 codes
(01)00884838110564(21)379(01)00884838110564(21)233(01)00884838110564(21)391(01)00884838110564(21)334(01)00884838110564(21)55(01)00884838110564(21)37(01)00884838110564(21)21(01)00884838110564(21)496(01)00884838110564(21)278(01)00884838110564(21)467(01)00884838110564(21)390(01)00884838110564(21)27(01)00884838110564(21)366(01)00884838110564(21)140(01)00884838110564(21)60(01)00884838110564(21)487(01)00884838110564(21)341(01)00884838110564(21)338(01)00884838110564(21)108(01)00884838110564(21)24(01)00884838110564(21)306(01)00884838110564(21)328(01)00884838110564(21)129(01)00884838110564(21)417(01)00884838110564(21)457(01)00884838110564(21)521(01)00884838110564(21)356(01)00884838110564(21)428(01)00884838110564(21)66(01)00884838110564(21)145(01)00884838110564(21)401(01)00884838110564(21)402(01)00884838110564(21)488(01)00884838110564(21)444(01)00884838110564(21)431(01)00884838110564(21)115(01)00884838110564(21)446(01)00884838110564(21)17(01)00884838110564(21)517(01)00884838110564(21)12(01)00884838110564(21)65(01)00884838110564(21)125(01)00884838110564(21)339(01)00884838110564(21)351(01)00884838110564(21)337(01)00884838110564(21)364(01)00884838110564(21)210(01)00884838110564(21)188(01)00884838110564(21)203(01)00884838110564(21)476(01)00884838110564(21)419(01)00884838110564(21)542(01)00884838110564(21)295(01)00884838110564(21)236(01)00884838110564(21)307(01)00884838110564(21)350(01)00884838110564(21)304(01)00884838110564(21)342(01)00884838110564(21)410(01)00884838110564(21)38
- Serial Numbers
212 of 212 codes
3792333913345537214962784673902736614060487341338108243063281294174575213564286614540140248844443111544617517126512533935133736421018820347641954229523630735030434241038
- Amount recalled
- 465 units
This recall covers 5 products
The agency filed one notice per product. You are reading one of them; the other 4 are below.
Where it was sold
- Nationwide
Why it was recalled
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.