IQon Spectral CT IQon Spectral CT Upgrades
Philips Medical Systems
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: PHILIPS MEDICAL SYSTEMS · 222 Jacobs St · Cambridge, MA · 02141-2289
Is this the product you have?
- IQon Spectral CT IQon Spectral CT Upgrades
- Lot codes and dates
FCO72800812FCO72800814 728331 728332DI (01)00884838059542
- Amount recalled
- 42
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
- PR
Why it was recalled
The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.