Reported Jul 15, 2026 · Ongoing · Updated Sep 16, 2026

OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

Integra LifeSciences Corp. (NeuroSciences)

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Integra LifeSciences Corp. (NeuroSciences) · 1100 Campus Rd · Princeton, NJ · 08540-6650

Is this the product you have?

  • OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.
Lot codes and dates

633 of 633 products

Products covered by this recall
Model / numberBarcodesProduct as published
303210381780268468UDI 10381780268468
305510381780268505UDI 10381780268505
306910381780268512UDI 10381780268512
307610381780268529UDI 10381780268529
308610381780268536UDI 10381780268536
309210381780268550UDI 10381780268550
309310381780268567UDI 10381780268567
309610381780268581UDI 10381780268581
309810381780268598UDI 10381780268598
309910381780268604UDI 10381780268604
313310381780268680UDI 10381780268680
313710381780268697UDI 10381780268697
314210381780268710UDI 10381780268710
314610381780268734UDI 10381780268734
314810381780268758UDI 10381780268758
317610381780268826UDI 10381780268826
317710381780268833UDI 10381780268833
318510381780268857UDI 10381780268857
320110381780268871UDI 10381780268871
320310381780268888UDI 10381780268888
Amount recalled
633 units

Where it was sold

  • Nationwide
  • PR

Why it was recalled

The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Integra LifeSciences Corp. (NeuroSciences) recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (Enforcement Report)

Read the FDA (weekly enforcement report) notice

Data as of Sep 19, 5:51 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.