Reported Jul 22, 2026 · Ongoing · Updated Sep 16, 2026

da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15

Intuitive Surgical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Intuitive Surgical, Inc. · 1266 Kifer Rd · Sunnyvale, CA · 94086-5304

Is this the product you have?

  • da Vinci SP Instrument Arm Drape, SP, 1-PACK
  • Part Number
    • 430013-13
    • 430013-15
Lot codes and dates
  • UDI-DI00886874113486
  • Lot
    • 3074LA0500
    • 1124LA0200
    • 2124LA0400
    • 1134LA0700
    • 1144LA0500
    • 7154LA0200
    • 6164LA0800
    • 1084LA0300
    • 3164LAA800
    • 4084LA0500
    • 4124LA0600
    • 1274LA0500
    • 2294LA0200
    • 1274LA0400
    • 4284LA0800
    • 6324LA0200
    • 1324LA0300
    • 4154LA0400
    • 2284LA0100
    • 4144LA1300
    • 3324LA0500
    • 2354LA0500
    • 7144LA0500
    • 4294LA0700
    • 3374LA0400
    • 5264LA0700
    • 3224LA0500
    • 3424LA0200
    • 1394LA0100
Amount recalled
11113

Where it was sold

Why it was recalled

Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a section of instrument arm, which may result in procedure delay or conversion to an alternate minimally invasive approach. Retrospectively reported recall.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Intuitive Surgical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (Enforcement Report)

Read the FDA (weekly enforcement report) notice

Data as of Sep 19, 7:10 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.