Reported Jul 22, 2026 · Ongoing · Updated Sep 16, 2026

Optima XR220amx, Mobile Digital-Ready Radiographic System

GE Medical Systems, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Medical Systems, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

  • Optima XR220amx, Mobile Digital-Ready Radiographic System
Lot codes and dates
  • UDI-DI00840682115261
  • System IDs
    • 05249MOB02.OLD
    • 205820DP1
    • 228818XR220
    • 229276M1
    • 240566P1
    • 301295OPTIMA2
    • 303788SBOPT1
    • 314251P5
    • 352LEEDP1
    • 386FLGDPXR3
    • 417347DPORT1
    • 512SAMCSB
    • 516632OPTIMA2
    • 540981ODP7
    • 563264OPT
    • 601445DP1
    • 615338HDP01
    • 661949DP2
    • 713441OPCDX1
    • 717657OPT1
    • 718206OPT240A
    • 727848SB1
    • 732321DP1
    • 757736DP3
    • 813302OPT1
    • 828213DXP1
    • 828286OPT1
    • 843847220D
    • 850070941
    • 865545DP3
    • 915821OPT1
    • 941342OP221
    • 972566XR220E
    • 985447TM4
    • SA1208MM05
    • SB2201700085WK
    • XR1034798WK7
    • XR1037005WK4
Amount recalled
38 units

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Medical Systems, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (Enforcement Report)

Read the FDA (weekly enforcement report) notice

Data as of Sep 19, 5:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.