Reported Jul 22, 2026 · Ongoing · Updated Sep 16, 2026
Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Daavlin Distributing Company
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Daavlin Distributing Company · 205 W Bement St · Bryan, OH · 43506-1264
Is this the product you have?
- Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
- Lot codes and dates
- Serial Numbers
212 of 212 codes
6-09278NB8CL6-09271NB8CL6-09055NB8CL6-09113NB8CL6-09204NB8CL6-09207NB8CL6-09116NB8CL6-09263NB8CL6-09272NB8CL6-09118NB8CL6-09275NB8CL6-09264NB8CL6-09060NB8CL6-09122NB8CL6-09216NB8CL6-09125NB8CL6-09353NB8CL6-09346NB8CL6-09128NB8CL6-09129NB8CL6-09133NB8CL6-09134NB8CL6-09266NB8CL6-09215NB8CL6-09355NB8CL6-09137NB8CL6-09359NB8CL6-09343NB8CL6-09140NB8CL6-09142NB8CL6-09143NB8CL6-09273NB8CL6-09065NB8CL6-09144NB8CL6-09237NB8CL6-09225NB8CL6-09145NB8CL6-09147NB8CL6-09148NB8CL6-09378NB8CL6-09252NB8CL6-09051NB8CL6-09260NB8CL6-09250NB8CL6-09149NB8CL6-09238NB8CL6-09212NB8CL6-09277NB8CL6-09048NB8CL6-09274NB8CL6-09259NB8CL6-09224NB8CL6-09057NB8CL6-09436NB12CL6-09438NB8CL6-09434NB12CL6-09270NB8CL6-09115NB8CL6-09111NB8CL6-09435NB12CL
- Serial Numbers
- Amount recalled
- 212 units
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
Why it was recalled
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Daavlin Distributing Company recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.