Reported Sep 2, 2026 · Ongoing · Updated Sep 16, 2026
ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600
IMRIS Imaging Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: IMRIS Imaging Inc · 1230 Chaska Creek Way Ste 100 · Chaska, MN · 55318-2707
Is this the product you have?
- ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600
- Lot codes and dates
- 114093-000 -
- UDI-DI00857534006356
- Serial Numbers
100056951000299810002997100054561000561210004372100056131000284110011119114093-600 -
- UDI-DI00857534006462
- Serial Numbers
10002996100045151000741110007376
- Amount recalled
- 13 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new IMRIS Imaging Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.