Reported Sep 16, 2026 · Ongoing

Getinge PVC Thoracic Catheter, Firm PVC Straight, 10 per Case; Size(Model Number): 1. 28FR(15028), 2. 36FR(15036);

Atrium Medical Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Atrium Medical Corporation · 40 Continental Blvd · Merrimack, NH · 03054-4332

Is this the product you have?

  • Getinge PVC Thoracic Catheter, Firm PVC Straight, 10 per Case
  • Size(Model Number)
    • 1. 28FR(15028)
    • 2. 36FR(15036)
Lot codes and dates
  • 1
  • Model Number(Size)15028(28FR)
  • UDI-DI00650862150285
  • Lot Number
    • ME244891
    • 3
  • Model Number(Size)15036(36FR)
  • UDI-DI00650862150360
  • Lot NumberME244850
Amount recalled
256

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

An Out of Specification result (OOS) related to bacterial endotoxin testing in a single lot of catheters.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Atrium Medical Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (Enforcement Report)

Read the FDA (weekly enforcement report) notice