Reported Sep 23, 2026 · Ongoing · Updated Oct 7, 2026

Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000

Datex-Ohmeda, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Datex-Ohmeda, Inc. · 3030 Ohmeda Dr · Madison, WI · 53718-6704

Is this the product you have?

  • Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000
Lot codes and dates
  • UDI-DI00840682102292
  • Serial Numbers

    475 of 475 codes

    • APWC01722
    • APWF00169
    • APWF00170
    • APWF00173
    • APWF00174
    • APWF00111
    • APWF00368
    • APWF00402
    • APWF00465
    • APWF00467
    • APWF00468
    • APWF00469
    • APWF00470
    • APWF00471
    • APWF00472
    • APWF00473
    • APWF00474
    • APWF00475
    • APWF00476
    • APWF00477
    • APWT00394
    • APWC01107
    • APWF00436
    • APWE01910
    • APWE01911
    • APWV00635
    • APWW01464
    • APWW01625
    • APWF00380
    • APWF00381
    • APWF00382
    • APWY00358
    • APWF00359
    • APWF00313
    • APWE01925
    • APWV02342
    • APWE01915
    • APWF00458
    • APWF00433
    • APWD02646
    • APWC01330
    • APWC01339
    • APWC01340
    • APWF00279
    • APWF00280
    • APWF00281
    • APWF00282
    • APWF00283
    • APWF00284
    • APWF00285
    • APWF00286
    • APWF00287
    • APWF00288
    • APWF00289
    • APWF00290
    • APWF00262
    • APWF00312
    • APWX00800
    • APWX00804
    • APWF00138
Amount recalled
475 units

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Datex-Ohmeda, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (Enforcement Report)

Read the FDA (weekly enforcement report) notice

Data as of Nov 4, 5:40 AM UTC