Reported Jun 25, 2014
Roche MODULAR ANALYTICS EVO Serum Work Area (SWA), Operator's Manual Software Version 08-02; Filter label: Filter (N)…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand · Product · Product details — Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs…
2 more products in this recall
FDA (openFDA device enforcement)
Reported May 21, 2014
ACCU-CHEK Compact Plus Test Strips For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip 51 count retail…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand · Reason for recall — Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing…
FDA (openFDA device enforcement)
Reported Apr 16, 2014
Cobas c8000 Modular Analyzer Series Cobas c502, Part Number 5964067001, a clinical chemistry analyzer intended for the…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand — The Initial Cassette Volume Check (ICVC) is not activated on the cobas c 502 module of the cobas 8000 modular analyzer series when manually filled…
FDA (openFDA device enforcement)
Reported Apr 9, 2014
Omni Micro-electrode/reference electrode for cobas b221 analyzer, Model/Catalog/Part Number: 03111873180 as a part of…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand · Product details — Reference electrode used beyond the guaranteed in-use 52 week lifetime,may leak and potentially cause erroneous pH and/or sodium results. Other Ion…
FDA (openFDA device enforcement)
Reported Dec 25, 2013
MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic acids from…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand — This issue is only relevant for customers importing xml files into the MagNA Pure 96 IVD instrument software for sample identification. There is a…
FDA (openFDA device enforcement)
Reported Dec 4, 2013
ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)\ The ACCU-CHEK Combo System…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand — There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK…
FDA (openFDA device enforcement)
Reported Nov 13, 2013
COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001,…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand · Reason for recall — Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE…
FDA (openFDA device enforcement)
Reported Sep 4, 2013
ACCU-CHECK FastClix Lancing Device. It is intended for use in conjunction with a blood glucose monitoring system.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand · Reason for recall — Internal Roche inspections have revealed that after the lancet is used for a finger stick, the needle may not fully retract back within the lancet…
FDA (openFDA device enforcement)
Reported Jul 24, 2013
B-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitative determination…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand · Reason for recall — Roche Diagnostics internal investigations in R&D have shown that the claims for biotin interference and the sample stability given in the Elecsys…
FDA (openFDA device enforcement)
Reported Jul 3, 2013
CoaguChek XS System (PST) The CoaguChek XS PT test strips are part of the CoaguChek XS System. The CoaguChek XS System…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand · Reason for recall — Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters.…
3 more products in this recall
FDA (openFDA device enforcement)
Reported May 29, 2013
Cobas¿ 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the cobas e 602 module only.…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand — When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause…
FDA (openFDA device enforcement)
Reported Apr 24, 2013
ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand · Reason for recall — Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent…
1 more product in this recall
FDA (openFDA device enforcement)
Reported Apr 10, 2013
Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with the cobas¿…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand · Product · Product details — Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to…
FDA (openFDA device enforcement)
Reported Jan 30, 2013
Roche COBAS INTEGRA 400 and 400 plus Analyzer, an in vitro diagnostic analyzer.
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.Roche Diagnostics Operations, Inc.
Matched: Brand · Product · Product details · Reason for recall — A software security issue with Oracles TNS-Listener component has been identified. TNS-Listener is supplied with the Roche COBAS INTEGRA 400/400 plus…
FDA (openFDA device enforcement)
Reported Dec 19, 2012
Cobas b 123 POC system cobas b 123 POC system is fully automated POC system for whole blood in vitro measurement of pH,…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand — It has been determined that cobas b 123 Fluid Pack COOX cuvettes for lot numbers 21426121 and 21426171, exhibit a significant increase in the…
FDA (openFDA device enforcement)
Reported Dec 12, 2012
Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001. The Cobas 8000 is a fully automated system…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Roche Diagnostics Operations, Inc.
Matched: Brand — On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent…
FDA (openFDA device enforcement)