Recall archive
195 recalls · page 3 of 9
Reported Jan 5, 2022
Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — ¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a…
FDA (openFDA device enforcement)
Reported Dec 8, 2021
Dimension Vista 1500 In-vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting,…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista Assays
1 more product in this recall
FDA (openFDA device enforcement)
Reported Nov 10, 2021
Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW)…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to…
FDA (openFDA device enforcement)
Reported Nov 3, 2021
Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11097543
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand — The firm has confirmed the potential for Atellica CH Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E…
1 more product in this recall
FDA (openFDA device enforcement)
Reported Jul 14, 2021
The ADVIA Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723034, is an automated, clinical chemistry…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the…
3 more products in this recall
FDA (openFDA device enforcement)
Reported Jun 23, 2021
Atellica CH 930 Analyzer with Photometer Lamp (SMN 11075676) Siemens Material Number: 11067000
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results
FDA (openFDA device enforcement)
Reported Apr 28, 2021
BCS XP System: Multipurpose system for in vitro coagulation studies. Product code: 10459330, 10461894, 10470625 -…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Reason for recall — Siemens Healthcare Diagnostics Products GmbH has confirmed that the BCS XP System may be affected by a potential Emicizumab carryover on patient…
FDA (openFDA device enforcement)
Reported Apr 21, 2021
ADVIA 2120 with Single Aspirate Autosampler (SMN 10316162) - US, automated hematology analyzer Software Versions 6.10…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand — Potential Sample Identification (SID) Mismatch with14-Character Barcodes
19 more products in this recall
FDA (openFDA device enforcement)
Reported Mar 24, 2021
Enzyme linked immunosorbent assay, for herpes simplex virus, hsv-2
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Reason for recall — Potential false positive HSV results. Siemens Healthcare Diagnostics, Inc. initiated a correction for ADVIA Centaur4P Herpes-2 lgG (HSV2) Onboard…
FDA (openFDA device enforcement)
Reported Mar 17, 2021
Dimension Vista HDLC - colorimetric method, lipoproteins - Product Usage: used as an aid in the diagnosis of lipid…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details · Reason for recall — Siemens Healthcare Diagnostics Inc. has confirmed the potential for erroneous results due to discolored reagent observed in some wells 5 and 6 of…
FDA (openFDA device enforcement)
Reported Jan 20, 2021
Atellica CH 930 Analyzer - In vitro diagnostic testing of clinical specimens. Product Siemens Material Number (SMN):…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient results
2 more products in this recall
FDA (openFDA device enforcement)
Reported Dec 30, 2020
ADVIA Gamma-Glutamyl Transferase (GGT) Reagent - Product Usage: For in vitro diagnostic use in the quantitative…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Gamma-Glutamyl Transferase (GGT) Reagent Low End Repeatability-Imprecision. Reagent may demonstrate a coefficient of variation (%CV) outside of the…
1 more product in this recall
FDA (openFDA device enforcement)
Reported Nov 25, 2020
Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new…
1 more product in this recall
FDA (openFDA device enforcement)
Reported Oct 28, 2020
Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro diagnostic test for the…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was…
FDA (openFDA device enforcement)
Reported Oct 21, 2020
ADVIA Chemistry Systems-Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to…
1 more product in this recall
FDA (openFDA device enforcement)
Reported Oct 14, 2020
Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient…
FDA (openFDA device enforcement)
Reported Sep 9, 2020
Dimension Vista High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Dimension Vista High Sensitivity Troponin I (TNIH) may exhibit a negative bias with patient samples
FDA (openFDA device enforcement)
Reported Aug 12, 2020
DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand — If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the…
FDA (openFDA device enforcement)
Reported Jul 8, 2020
(1) CentraLink Data Management System software version 16.0.2 Siemens Material Number (SMN): 11313246 (2) CentraLink…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of…
1 more product in this recall
FDA (openFDA device enforcement)
Reported Jul 1, 2020
Atellica Solution IM1300 Analyzer, Siemens Material Number 11066001, UDI # 00630414007960
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Insufficient sample may not be detected when using Tube Top Sample Cup (TTSC) or V-shaped Sample Cup sample containers with the Atellica IM1300 and…
3 more products in this recall
FDA (openFDA device enforcement)
Reported Jun 17, 2020
ADVIA Chemistry Fructosamine Reagent (FRUC), Siemens Material Number 10361941, UDI Number 00630414558073 - Product…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing…
5 more products in this recall
FDA (openFDA device enforcement)
Reported May 6, 2020
Atellica CH 930 Analyzer, Siemens Material Number 11067000 with software version v1.21.0 SP1 or lower
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the…
FDA (openFDA device enforcement)
Reported Apr 1, 2020
Atellica CH Direct Bilirubin_2 assay In vitro Diagnostic for the determination of direct bilirubin SMN: 11097532 -…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
6 more products in this recall
FDA (openFDA device enforcement)
Reported Mar 25, 2020
IMMULITE 1000 Systems E2 Estradiol Catalog #: US: LKE21(D) and OUS: LKE21 - Product Usage: For in vitro diagnostic use…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Healthcare Diagnostics, Inc.
Matched: Brand · Product · Product details — Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician…
1 more product in this recall
FDA (openFDA device enforcement)