Back to recall look-up

Recall archive

94,355 stored notices, from Jun 20, 2012 to Sep 23, 2026.

7 recalls · page 1 of 1

Medical devices

Reported Apr 27, 2022

Philips Respironics V60 Ventilator (All Models, including: 1053613, 1053614, 1053615, 1053616, 1053617, 1053618,…

High riskUsing this product could cause serious injury, illness or death — stop using it now.

Respironics California, LLC

Matched: Brand · Product · Product detailsVentilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator…

2 more products in this recall

FDA (openFDA device enforcement)

Medical devices

Reported Mar 9, 2022

Philips Respironics V60 Ventilator Part Number 1053617

High riskUsing this product could cause serious injury, illness or death — stop using it now.

Respironics California, LLC

Matched: Brand · Product · Product detailsA limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the…

1 more product in this recall

FDA (openFDA device enforcement)

Medical devices

Reported Jul 28, 2021

Philips Respironics V60 Ventilator with Software Versions 3.00 and 3.10. The Respironics V60/V60 Plus Ventilator is an…

High riskUsing this product could cause serious injury, illness or death — stop using it now.

Respironics California, LLC

Matched: Brand · Product · Product detailsIt has been identified that ventilators equipped with High Flow Therapy (Software Versions 3.00 and 3.10), that includes a designed safety mechanism…

1 more product in this recall

FDA (openFDA device enforcement)

Medical devices

Reported Mar 3, 2021

PHILIPS BATTERY, V60 14.4V, 11.0 Ah, 163 Wh, REF 1076374(1058272), LOT M94663-P1 - Product Usage: The ventilator is…

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

Respironics California, LLC

Matched: Brand · Product detailsDue to unqualified "design verification" batteries being shipped to customers.

FDA (openFDA device enforcement)

Medical devices

Reported Feb 3, 2021

Philips Respironics V60 Ventilator Part Number 1053613 - Product Usage: intended to be used only with various…

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

Respironics California, LLC

Matched: Brand · Product · Product detailsVentilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator…

32 more products in this recall

FDA (openFDA device enforcement)

Medical devices

Reported Apr 8, 2020

Philips V60 Ventilators with Power Management PCBA part number 1055906

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

Respironics California, LLC

Matched: BrandSolder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and…

FDA (openFDA device enforcement)

Medical devices

Reported Feb 19, 2020

Respironics V60 Ventilator, Service # 850008, REF: 1053613, 1053614, 1053617, 1053614

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

Respironics California, LLC

Matched: Brand · Product · Product detailsVentilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning…

3 more products in this recall

FDA (openFDA device enforcement)

RecallNow is not a government agency. Every notice above is published by the FDA, USDA FSIS, or CPSC and links back to its source. Back to the look-up