Reported Oct 16, 2024 · Ongoing

Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390,…

Philips North America

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips North America · 222 Jacobs St · Cambridge, MA · 02141-2296

Is this the product you have?

    • Patient Information Center iX
    • Product Code 866386
    • and Patient Information Center iX Expand
    • Product Code 866390
    • Software Revision 4.0.1
Lot codes and dates
  • UDI-DI00884838104594
  • Serial Numbers

    591 of 591 codes

    • 0Y5Z-7FHX-N
    • 5H20-1DNN-E
    • 1762-74NU-5
    • 1C0V-08NR-0
    • 6U53-30PD-1
    • 6M5J-0JLM-H
    • 2M2G-15LB-E
    • 4U3A-3JGK-K
    • 0439-16GM-G
    • 3A1T-22HW-U
    • 534U-57JE-Y
    • 0D6U-5ZH1-D
    • 3H5C-6JK9-X
    • 3K0C-3EH3-H
    • 3Z47-12JD-V
    • 524N-43JR-2
    • 5N40-2RHR-N
    • 1331-5BNT-B
    • 1V4F-34MB-G
    • 1B4R-37N1-3
    • 6D3L-28MC-M
    • 7Y52-29H4-L
    • 7Y5Z-7FHX-F
    • 1X6W-21KC-W
    • 6278-09GU-Y
    • 2C6N-3KLU-C
    • 3E5Y-32PN-4
    • 5T2L-2TP9-V
    • 021Z-1WJ2-5
    • 1W6W-21KC-Z
    • 7L3H-46MR-L
    • 194D-2YHB-T
    • 1Y7A-49M4-6
    • 2Z3C-1AHM-3
    • 4T51-53NK-A
    • 1P4A-44NB-V
    • 3B5M-55J9-5
    • 5V57-0HN3-J
    • 7V5Z-7FHX-R
    • 224B-2HJF-F
    • 5G1G-01JC-Z
    • 6Y7F-06MT-0
    • 676T-6XMZ-M
    • 5F1G-01JC-2
    • 5K2B-6DHP-4
    • 5T45-7ELE-C
    • 0V7B-4EJH-A
    • 1F23-36KM-0
    • 2C50-6KKD-R
    • 3107-4RG3-6
    • 3X5B-1NH8-2
    • 6U32-24GX-0
    • 5K2L-2TP9-D
    • 4T2R-6HHJ-D
    • 613L-28MC-T
    • 7K21-0XNW-C
    • 062U-1VNP-B
    • 5M78-09GU-6
    • 6226-14LG-X
    • 0V6R-0PLF-7
Amount recalled
591 devices

Where it was sold

  • Nationwide

Why it was recalled

During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips North America recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:55 PM UTC