Reported Oct 7, 2026 · Ongoing · Updated Oct 13, 2026

Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T…

Philips North America

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips North America · 222 Jacobs St · Cambridge, MA · 02141-2296

Is this the product you have?

  • Ingenia 3.0T and 3.0T CX
  • Model number (REF)
    • 1. 781342 (3.0T)
    • 2. 781377 (3.0T)
    • 3. 782103 (3.0T)
    • 4. 781271 (3.0T CX)
  • Software Versions
    • R11
    • R12 and R13
Lot codes and dates

5 of 5 products

Products covered by this recall
#BarcodesProduct as published
1Model number (REF)
200884838055339781342 (3.0T); UDI-DI: 00884838055339; Serial Numbers: 71001, 71002, 71008, 71031, 71038, 71051, 71061, 71072, 71084, 71088, 71089, 71095, 71101, 71154, 71163, 71164, 71166, 71184, 71185, 71193, 71210, 71231, 71269, 71271, 71288, 71290, 71297, 71301, 71303, 71309, 71311, 71312, 71322, 71335, 71336, 71340, 71341, 71344, 71355, 71375, 71385, 71393, 71412, 71413, 71420, 71422, 71427, 71438, 71441, 71443, 71445, 71446, 71449, 71450, 71459, 71462, 71463, 71475, 71486, 71487, 71489, 71491, 71513, 71521, 71528, 71556, 71558, 71562, 71589, 71604, 71607, 71612, 71615, 71619, 71620, 71639, 71641, 71646, 71665, 71673, 71688, 71697, 71704, 71710, 71715, 71729, 71730, 71739, 71754, 71756, 71757, 71759, 71813, 71817, 71828, 71831, 71841, 71849, 71850, 71853, 71862, 71885, 71896, 71921, 71923, 71954, 71959, 71960, 71962, 71963, 71964, 71986, 71992, 85006, 85020, 85032, 85038, 85046, 85066, 85068, 85076, 85078, 85085, 85087, 85091, 85092, 85102, 85126, 85148, 85150, 85154, 85191, 85201, 85216, 85245, 85279, 85377
300884838009813Model Number (REF): 781377 (3.0T); UDI-DI: 00884838009813; Serial Numbers: 42012, 42014, 42015, 42017, 42033, 42036, 42048, 42059, 42088, 42092, 42109, 42125, 42126, 42127, 42146, 42169, 42196, 42198, 42206, 42210, 42218, 42220, 42224, 42228, 42234, 42238, 42247, 42258, 42265, 42271, 42282, 42288, 42293, 42305, 42307, 42321, 42322, 42330, 42342, 42352, 42353, 42359, 42360, 42386, 42389, 42396, 42403, 42409, 42412, 42416, 42425
400884838098299Model Number (REF): 782103 (3.0T); UDI-DI: 00884838098299; Serial Numbers: 85646, 85652, 85679, 85686, 85708, 85723, 85758, 85759, 85762
500884838068452Model Number (REF): 781271 (3.0T CX); UDI-DI: 00884838068452; Serial Numbers: 78008, 78009, 78011, 78012, 78017, 78018, 78027, 78028, 78033, 78035, 78036, 78039, 78040, 78042, 78045, 78053, 78060, 78062, 78063, 78067, 78068, 78070, 78095, 78100, 78103, 78112, 78113, 78115, 78122, 78145, 78147, 78148, 78155, 78166, 78167, 78186, 78195, 78201, 78202, 78208, 78215, 78217, 78229, 78277, 78286, 78287
Amount recalled
243

This recall covers 6 products

The agency filed one notice per product. You are reading one of them; the other 5 are below.

Where it was sold

  • Nationwide
  • GA

Why it was recalled

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips North America recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Nov 4, 1:03 AM UTC