Reported Oct 11, 2023 · Ongoing

4500MD Triple Quad Mass Spectrometer, Part Number 5032522

AB Sciex, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: AB Sciex, LLC · 500 Old Connecticut Path · Framingham, MA · 01701-4574

Is this the product you have?

    • 4500MD Triple Quad Mass Spectrometer
    • Part Number 5032522
Lot codes and dates
  • UDI-DI00842365102049
  • Serial Numbers
    • EJ230051806PL
    • EJ230061807PL
    • EJ230071808
    • EJ230101810
    • EJ230111811
    • EJ230121811
    • EJ230141811
    • EJ230151812
    • EJ230161812
    • EJ230171901
    • EJ230181901
    • EJ230191901
    • EJ230211905
    • EJ230221908
    • EJ230231908
    • EJ230251909
    • EJ230271910
    • EJ230281910
    • EJ230311911
    • EJ230331912
    • EJ230362003
    • EJ230372003
    • EJ230392004
    • EJ230412009
    • EJ230422011
    • EJ230442012
    • EJ230452012
    • EJ230472101
    • EJ230482103
    • EJ230512105
    • EJ240012110
    • EJ240022111
    • EJ240052111
    • EJ240082207
    • EJ240092209
    • EJ240142307
    • EJ240132305
    • EJ240062201
Amount recalled
38

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

  • Nationwide

Why it was recalled

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new AB Sciex, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:56 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.