Reported Oct 11, 2023 · Ongoing
Citrine Triple Quad Mass Spectrometer, Part Number 5063684
AB Sciex, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: AB Sciex, LLC · 500 Old Connecticut Path · Framingham, MA · 01701-4574
Is this the product you have?
- Citrine Triple Quad Mass Spectrometer, Part Number 5063684
- Lot codes and dates
- UDI-DI00842365100533
- Serial Numbers
DE220001803PLDE230011804DE230031806DE230041807DE230051808DE230071901DE230081903DE230091903DE230162003DE230272105DE230292105DE240002106DE240022111DE240042203
- Amount recalled
- 14
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
Why it was recalled
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new AB Sciex, LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.