Reported Oct 26, 2022 · Ongoing
Medicina Oral Tip Syringe 1ml Code: OT01
Medicina Uk Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medicina Uk Ltd · Unit 2 · Bolton, N/A · N/A
Is this the product you have?
- Medicina Oral Tip Syringe 1ml CodeOT01
- Lot codes and dates
- UDI-DI(01)05060278503424
- Batch number
HM18147HM19012HM19025HM19088HM19164HM20063 HM21015HM21072
- Amount recalled
- 29 boxes (2900 syringes)
This recall covers 8 products
The agency filed one notice per product. You are reading one of them; the other 7 are below.
Where it was sold
- Nationwide
Why it was recalled
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medicina Uk Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.