Reported Oct 26, 2022 · Ongoing
Medicina Reusable Oral syringe 5ml Code: OTH05
Medicina Uk Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medicina Uk Ltd · Unit 2 · Bolton, N/A · N/A
Is this the product you have?
- Medicina Reusable Oral syringe 5ml CodeOTH05
- Lot codes and dates
- UDI-DI(01)05056115403491 (01)05056115403491(17)260700(10)HM21055
- Batch number
HM18043HM20020HM21055
- Amount recalled
- 26 boxes (2600 syringes)
This recall covers 8 products
The agency filed one notice per product. You are reading one of them; the other 7 are below.
Where it was sold
- Nationwide
Why it was recalled
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification