Reported Nov 6, 2013 · Terminated
Aplio MX Diagnostic Ultrasound System SSA-780A Used for routine diagnostic imaging
Toshiba American Medical Systems Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Toshiba American Medical Systems Inc · 2441 Michelle Dr · Tustin, CA · 92780-7047
Is this the product you have?
- Aplio MX Diagnostic Ultrasound System SSA-780A Used for routine diagnostic imaging
- Lot codes and dates
- Lot Numbers not utilized
- Serial Numbers
427 of 427 codes
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
- Amount recalled
- 427 units (1352 total)
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
- PR
Why it was recalled
The firm initiated this recall of certain Toshiba Diagnostic Ultrasound Systems because it has been found that under certain conditions, the temperature of the transducer head may increase in 2D+PW mode or 2D+CDI (POWER, ADF) +PW mode. When the system is used under both conditions listed below, the temperature of the transducer head connected to the system may increase in temperature and possibly
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Toshiba American Medical Systems Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.