Reported Oct 21, 2026 · Ongoing · Updated Oct 26, 2026
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115
Steris Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Steris Corporation · 5960 Heisley Rd · Mentor, OH · 44060-1834
Is this the product you have?
- RAPICIDE PA Part A High-Level Disinfectant
- Model/Catalog NumberML020115
- Lot codes and dates
- UDI00724995245658
- Lot numbers with expiration dates687046 10/5/2027 688017 10/19/2027
- Amount recalled
- 374 units
Where it was sold
- Nationwide
Why it was recalled
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Steris Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.