Reported Dec 17, 2025 · Ongoing

AMSCO 7052HP Washer/Disinfector Model/Catalog Number: 7052HP Product Description: The AMSCO 7052HP Washer/Disinfector…

Steris Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Steris Corporation · 5960 Heisley Rd · Mentor, OH · 44060-1834

Is this the product you have?

  • AMSCO 7052HP
  • Washer/Disinfector Model/Catalog Number7052HP
  • Product DescriptionThe AMSCO 7052HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities
Lot codes and dates
  • Lot Code
    • UDI:00724995176587
    • 00724995176594
    • 00724995176570
  • Serial number of distributed units
    • 3619625011
    • 3621625009
    • 3621625012
    • 3621925006
    • 3621925009
    • 3622425006
    • 3622725007
    • 3623425005
    • 3623825002
    • 3624525002
Amount recalled
10

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Steris Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:54 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.