Reported Nov 29, 2017 · Completed
ARCHITECT c4000 part number 02P24; ARCHITECT c8000 part number 01G06; ARCHITECT c16000 part number 03L77 Clinical…
Abbott Laboratories, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Abbott Laboratories, Inc · 1921 Hurd Dr · Irving, TX · 75038-4313
Is this the product you have?
- ARCHITECT c4000 part number 02P24; ARCHITECT c8000 part number 01G06; ARCHITECT c16000 part number 03L77 Clinical Chemistry systems.
- Lot codes and dates
c400073c400103c400498c400600c400890c400934c400935c401018c401055c401056c401182c401505c401520c401686c401887c460196c460200c460943c461550c803669c803753c1600144c1600249c1600251c1600641c1600720c1600819c1600859c1600976c1601350c1601351
- Amount recalled
- 34 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Why it was recalled
The Peristaltic Head tubing placed on the ARCHITECT c4000, c8000, and c16000 has the possibility to leak due to a manufacturing issue associated with the tubing connector.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Abbott Laboratories, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.