Reported Oct 22, 2025 · Ongoing
Cardiac Workstation 7000; Model Number: 860441;
Philips North America
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips North America · 222 Jacobs St · Cambridge, MA · 02141-2296
Is this the product you have?
- Cardiac Workstation 7000
- Model Number860441
- Lot codes and dates
- Model Number860441
- UDI-DI00884838094833
- Serial Numbers
US42520284US32520248US32520244US22520238US22520221
- Amount recalled
- 29 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
This product could cause temporary or reversible health problems — stop using it.