Reported Nov 12, 2014 · Terminated

COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological

Iba Dosimetry GmbH

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Iba Dosimetry Gmbh · Bahnhofstr. 5 · Schwarzenbruck, N/A · N/A

Is this the product you have?

  • COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological
Lot codes and dates
  • N/A
Amount recalled
124 units

Why it was recalled

Error in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over or under-estimation of the pretreatment delivered dose.

This product could cause temporary or reversible health problems — stop using it.