Reported Dec 18, 2013 · Terminated

iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological

Iba Dosimetry GmbH

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Iba Dosimetry Gmbh · Bahnhofstr. 5 · Schwarzenbruck · N/A

Is this the product you have?

  • iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological
Lot codes and dates
  • SW Version 3.0. Catalong Number CS10-100
Amount recalled
33 units

Where it was sold

Why it was recalled

The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Iba Dosimetry GmbH recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 24, 2014 · Data as of Sep 19, 6:16 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.