Reported Dec 6, 2017 · Terminated

SIEMENS N Protein Standard SL, for use on the BN II, BN ProSpec Systems, REF/Product Code OQIM 13, Siemens Material…

CSL Behring GmbH

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: CSL Behring GmbH · Emil-von-Behring-Str. 76 · Marburg, N/A · N/A

Is this the product you have?

  • SIEMENS N Protein Standard SL, for use on the BN II, BN ProSpec Systems, REF/Product Code OQIM 13, Siemens Material Number (SMN)10446073
Lot codes and dates
  • Lot Numbers(UDI)
    • 083611A (00842768007590083611A20170711)
    • 083611B (00842768007590083611B20170711)
    • 083611C (00842768007590083611C20170711)
    • 083611D (00842768007590083611D20170711)
    • exp. date 2017-07-11
    • 083612 (0084276800759008361220171014)
    • 083612A (00842768007590083612A20171014)
    • 083612B (00842768007590083612B20171014)
    • exp. date 2017-10-14
    • 083614A (00842768007590083614A20180127)
    • 083614B (00842768007590083614B20180127)
    • 083614C (00842768007590083614C20180127)
    • exp. date 2018-01-27
    • 083615 (0084276800759008361520180411)
    • 083615A (00842768007590083615A20180411)
    • 083615B (00842768007590083615B20180411)
    • 083615C (00842768007590083615C20180411)
    • exp. date 2018-04-11
    • 083616B (00842768007590083616B20180730)
    • 083616C (00842768007590083616C20180730)
    • 083616D (00842768007590083616D20180730)
    • exp. date 2018-07-30
Amount recalled
N/A

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Why it was recalled

Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new CSL Behring GmbH recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 7, 2020 · Data as of Sep 19, 5:22 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.