Reported Dec 6, 2017 · Terminated
SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems
CSL Behring GmbH
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: CSL Behring GmbH · Emil-von-Behring-Str. 76 · Marburg, N/A · N/A
Is this the product you have?
SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems
3 of 3 products
| Model / number | Product as published |
|---|---|
OQIN19 | N/T Protein Control SL/L: REF/ , SMN 10446079 (Low) |
OQIO19 | N/T Protein Control SL/M, REF/ , SMN 10446085 (Mid) |
OQIP19 | N/T Protein Control SL/H, REF/ , SMN 10446089 (High) |
- Lot codes and dates
- SMN 10446079Lot (UDI): 084647A (00842768016127084647A20170913), 084647C (00842768016127084647C20170913) exp. date 2017-09-13; 084648(0084276801612708464820180430), 084648C (00842768016127084648C20180430), 084648H (00842768016127084648H20180430), exp. date 2018-04-30. ---
- SMN 10446085Lot (UDI): 084749D (00842768016134084749D20170917) exp. date 2017-09-17; 084750 (03 0084276801613408475020180203), 084750B (00842768016134084750B20180203), 084750E (00842768016134084750E20180203) exp. date 2018-02-03. ---
- SMN 10446089Lot (UDI) 084847A (00842768016110084847A20170613), 084847B (00842768016110084847B20170613) exp. date 2017-06-13; 084848A (00842768016110084848A20171217), 084848D (00842768016110084848D20171217), 084848E (00842768016110084848E20171217), 084848H (00842768016110084848H20171217), 084848L (00842768016110084848L20171217) exp. date 2017-12-17; 084849 (0084276801611008484920180704), 084849C (00842768016110084849C20180704), 084849D (00842768016110084849D20180704) 084849E (00842768016110084849E20180704) exp. date 2018-07-04
- Amount recalled
- N/A
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Why it was recalled
Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new CSL Behring GmbH recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.