Reported Dec 6, 2017 · Terminated

SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems

CSL Behring GmbH

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: CSL Behring GmbH · Emil-von-Behring-Str. 76 · Marburg, N/A · N/A

Is this the product you have?

SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems

3 of 3 products

Products covered by this recall
Model / numberProduct as published
OQIN19N/T Protein Control SL/L: REF/ , SMN 10446079 (Low)
OQIO19N/T Protein Control SL/M, REF/ , SMN 10446085 (Mid)
OQIP19N/T Protein Control SL/H, REF/ , SMN 10446089 (High)
Lot codes and dates
  • SMN 10446079Lot (UDI): 084647A (00842768016127084647A20170913), 084647C (00842768016127084647C20170913) exp. date 2017-09-13; 084648(0084276801612708464820180430), 084648C (00842768016127084648C20180430), 084648H (00842768016127084648H20180430), exp. date 2018-04-30. ---
  • SMN 10446085Lot (UDI): 084749D (00842768016134084749D20170917) exp. date 2017-09-17; 084750 (03 0084276801613408475020180203), 084750B (00842768016134084750B20180203), 084750E (00842768016134084750E20180203) exp. date 2018-02-03. ---
  • SMN 10446089Lot (UDI) 084847A (00842768016110084847A20170613), 084847B (00842768016110084847B20170613) exp. date 2017-06-13; 084848A (00842768016110084848A20171217), 084848D (00842768016110084848D20171217), 084848E (00842768016110084848E20171217), 084848H (00842768016110084848H20171217), 084848L (00842768016110084848L20171217) exp. date 2017-12-17; 084849 (0084276801611008484920180704), 084849C (00842768016110084849C20180704), 084849D (00842768016110084849D20180704) 084849E (00842768016110084849E20180704) exp. date 2018-07-04
Amount recalled
N/A

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Why it was recalled

Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new CSL Behring GmbH recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 7, 2020 · Data as of Sep 19, 4:44 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.